7 Test Sequence Application
308 Keysight U8903B User’s Guide
Prompt sub-step settings
The prompt sub-step inserts a prompt message with reply option to the measurement. This
provides you the information and opportunity to interact with the sequence. Press OK to
close the prompt window and continue the test sequence. The prompt sub-step settings
menu page is displayed as shown in Figure 7-13.
Figure 7-13 Prompt sub-step settings menu page
Table 7-12 Prompt sub-step settings menu description
Menu Description
Message Press the Message softkey to set the prompt message at the lower half of the main display.
Prompt Icon
Press the Prompt Icon softkey to select the prompt icon to be displayed.
–None
–Hand
Question
–Exclamation
Asterisk
Dialog Settings
Press the Dialog Settings softkey to add additional settings to the prompt window.
–Timeout
When selected, the timeout value sets the maximum prompt window display time, in seconds. If the timeout is reached,
the current measurement fails. When unselected, the prompt window will remain open until user closes it.
–Cancel
When selected, a Cancel button will be added to the prompt window. Press the Cancel button to close the prompt
window and stop the test sequence.
Timeout
Press the Timeout softkey to set the prompt timeout value in seconds. This setting is only available if Timeout is selected in
the dialog settings.
Minimum: 1 s
Maximum: 3600 s (1 hour)