302 User’s Manual TOS93 Series
Specifications | Leakage current test section
Voltmeter
accuracy
2
Range1 DC ±(2.0 % of reading + 1 mV)
RMS 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 1 mV)
15 Hz ≤ f ≤ 100 kHz ±(2.0 % of reading + 1 mV)
100 kHz < f ≤ 1 MHz
±(5.0 % of reading + 1 mV)
Peak 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 1 mV)
15 Hz ≤ f ≤ 1 kHz ±(5.0 % of reading + 1 mV)
1 kHz < f ≤ 100 kHz ±(5.0 % of reading + 1 mV)
100 kHz < f ≤ 1 MHz
±(15.0 % of reading + 1 mV)
Range2 DC ±(2.0 % of reading + 5 mV)
RMS 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 5 mV)
15 Hz ≤ f ≤ 100 kHz ±(2.0 % of reading + 4 mV)
100 kHz < f ≤ 1 MHz
±(5.0 % of reading + 5 mV)
Peak 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 5 mV)
15 Hz ≤ f ≤ 1 kHz ±(5.0 % of reading + 5 mV)
1 kHz < f ≤ 100 kHz ±(5.0 % of reading + 5 mV)
100 kHz < f ≤ 1 MHz
±(15.0 % of reading + 5 mV)
Range3 DC ±(2.0 % of reading + 25 mV)
RMS 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 10 mV)
15 Hz ≤ f ≤ 100 kHz ±(2.0 % of reading + 10 mV)
100 kHz < f ≤
1 MHz
±(5.0 % of reading + 10 mV)
Peak 0.1
Hz ≤ f < 15 Hz ±(5.0 % of reading + 25 mV)
15 Hz ≤ f ≤ 1 kHz ±(2.0 % of reading + 25 mV)
1 kHz < f ≤ 100 kHz ±(5.0 % of reading + 25 mV)
100 kHz < f ≤
1 MHz
±(15.0 % of reading + 25 mV)
Range4 DC ±(2.0 % of reading + 0.25 V)
RMS 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 0.1 V)
15 Hz ≤ f ≤ 100 kHz ±(2.0 % of reading + 0.1 V)
100 kHz < f ≤ 1 MHz
±(5.0 % of reading + 0.1 V)
Peak 0.1 Hz ≤ f < 15 Hz ±(5.0 % of reading + 0.25 V)
15 Hz ≤ f ≤ 1 kHz ±(2.0 % of reading + 0.25 V)
1 kHz < f ≤ 100 kHz ±(5.0 % of reading + 0.25 V)
100 kHz < f ≤ 1 MHz
±(15.0 % of reading + 0.25 V)
Item Specifications