86 User’s Manual TOS93 Series
Withstanding Voltage and Insulation Resistance Tests | Starting a Test
Voltage rise time display
While the voltage is rising, “RISE” is displayed along with the elapsed time. The remaining rise time is
displayed with a bar graph. When the time specified by Rise Time elapses, “TEST” is shown, and the test
starts.
Test time display
During the test time, “TEST” and the elapsed time are displayed. If Test Time is set to on, the remaining
test time is displayed with a bar graph. If Test Time is set to off, the elapsed time is displayed up to
3600000.0 seconds. If the elapsed time is greater than 3 600 000.0, the display is fixed at 3 600 000.0.
Voltage fall time display
If Fall Time is set to on, “FALL” and the elapsed time are displayed while the voltage is falling. The
remaining fall time is displayed with a bar graph.
Elapsed time. The remaining rise time is
displayed with a bar graph.
Elapsed time. If Test Time is set to on, the
remaining test time is displayed with a bar graph.
Elapsed time. The remaining fall time is
displayed with a bar graph.