Managing Special Waveform Functions (Conventional UT Mode) 101
5. Managing Special Waveform Functions
(Conventional UT Mode)
This chapter describes how to manage special waveform functions. The topics are as
follows:
• Reject
• Peak Memory
• Peak Hold
• Freeze
• Grid Modes
5.1 Reject
The 1/5 > Receiver > Reject parameter allows eliminating unwanted, low-level signals
from the display. The reject function is linear and adjustable from 0
% to 80 % FSH.
Increasing the reject level does not affect the amplitude of the signals above the reject
level.
The reject function can also be used in the unrectified 1/5 > Receiver > Rect = RF
mode.
The reject level is displayed as an horizontal line on the instrument display (see
Figure 5-1 on page 102) or two lines in the case of the 1/5 > Receiver > Rect = RF mode.