910-269-EN, Rev. B, June 2011
148 Chapter 9
Figure 9-8 The Enter Value for Zero Cal dialog box
11. Adjust the value to match the known thickness of the gated indication (0.100 in. in
this example), and then choose Continue to continue to the second calibration
step.
If for any reason you need to exit without acquiring calibration data, press the
[RIGHT] or [LEFT] arrow key to highlight the Cancel box, and then press
[CHECK].