910-269-EN, Rev. B, June 2011
Software Features and Options (Conventional UT Mode) 227
Figure 11-22 Gain curve adjusted DGS
11.3.5 Registration Level Adjustment
The registration level of the DGS/AVG curve defines the height of the main curve. The
curve height represents the amplitude from a flat-bottom hole with a diameter of the
registration level at different depths. This is usually equal to the critical flaw size for
the application. The EPOCH
1000 Series allows you to adjust this registration level
during a live inspection.
This curve-height adjustment is possible because the DGS/AVG curves are
calculated based on a captured reference-reflector and mathematical probe data.
This allows the EPOCH 1000 Series to plot the attenuation curve (in steel) for a
particular size reflector without having to acquire individual data points, as is
required in a DAC/TVG setup. This is one of the key advantages of the DGS/AVG
sizing technique over the DAC/TVG sizing technique.