910-269-EN, Rev. B, June 2011
158 Chapter 9
Figure 9-15 Example of a gated calibration signal
6. Once a steady reading is achieved, choose 1/5 > Auto CAL > CAL Zero.
The screen freezes and the Enter Value for Zero Cal dialog box appears.
Figure 9-16 The Enter Value for Zero Cal dialog box
7. Adjust the value to match the known thickness of the gated indication (4.000 in. in
this example), and then choose Continue to continue to the second calibration
step.
Thickness measurement