910-269-EN, Rev. B, June 2011
Probe and Beam Setup (Phased Array Mode) 241
Figure 12-2 S-scan image 40°–70° with 4 in. sound path
The Beam setup page has three main sections of items that must be adjusted to ensure
proper phased array setup and calibration.
12.2.1 Probe and Wedge Selection
The first main section of items in the Beam setup page includes Probe ID and Wedge
ID. These are the two most important items to adjust to ensure proper phased array
probe setup.
Probe ID
The name of the phased array probe connected to the instrument. This item is
automatically populated with the correct probe ID name in most cases when
using an Olympus phased array probe.
Wedge ID
The name of the wedge attached to the phased array probe. Changing the Wedge
ID adjusts many of the remaining items in the Beam setup page to common
settings based on the type of wedge selected.