910-269-EN, Rev. B, June 2011
292 Chapter 17
9. Once a steady reading is achieved, choose 1/5 > PA Cal > Depth 1.
The screen freezes and the Enter Value For Thin Standard dialog box appears.
Figure 17-14 The Enter Value For Thin Standard dialog box
10. Adjust the value to match the known thickness of the gated indication (4.000 in. in
this example), and then choose Continue to continue to the second calibration
step.
If for any reason you need to exit without acquiring calibration data, use the
[LEFT] or [RIGHT] arrow key to highlight Cancel, and then press [CHECK].