910-269-EN, Rev. B, June 2011
Managing Special Waveform Functions (Conventional UT Mode) 103
Figure 5-2 Peak memory signal envelope example
This function is useful when it is necessary to find the peak from an indication during
an angle beam inspection.
The peak memory function cannot be activated in the unrectified 1/5 > Receiver >
Rect = RF mode.
To activate peak memory function
1. Press [PEAK MEM].
The symbol appears in the flag area to indicate that the function is active.
2. Scan over a reflector to acquire the echo envelope.
3. Press [PEAK MEM] again to turn off the peak memory function.
Peak memory flag
Peak memory signal
envelope (green line)
Current live waveform (white
line)