The analog output setup parameters are located in the A-Out setup page accessible by
choosing 3/5
> Meas Setup > A-Out (see Figure 8-2 on page 129).
Figure 8-2 The A-Out setup page
There are four main parameters that control each analog output signal:
Reading<n>
Used to select the measurement (thickness or amplitude) to output to the A
NALOG
O
UT connector.
Output<n>
Used to select the range of voltage output from the unit (0-1 V or 0-10 V).
Mapping<n>
Used to select the scale of the output voltage based either on the full-screen range
(Range) or on the gate width (GateWidth).
Load<n>
Used to select the value for the impedance of the peripheral device measuring the
analog output of the EPOCH 1000 Series instrument.