910-269-EN, Rev. B, June 2011
Calibrating the EPOCH 1000 Series (Phased Array Mode) 281
Figure 17-3 Example of the second gated calibration signal
12. Once a steady reading is achieved, choose 1/5 > PA CAL > Depth 2.
The screen freezes and the Enter Value for Thick Standard dialog box appears.
Figure 17-4 The Enter Value for Thick Standard dialog box
13. Adjust the value to match the known thickness of the gated indication (0.500 in. in
this example), and then choose Continue to continue to complete the calibration
process.