EasyManua.ls Logo

Olympus EPOCH 1000 Series - Figure 17-9 Example of the Captured Echo

Olympus EPOCH 1000 Series
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
910-269-EN, Rev. B, June 2011
Calibrating the EPOCH 1000 Series (Phased Array Mode) 287
Figure 17-9 Example of the captured echo
7. Choose 1/5 > PA CAL > Add to begin the calibration process.
The A-scan at the top of the screen is replaced by a new view representing the
amplitude reading (yellow line) for the gated indication across all focal laws (see
Figure 17-7 on page 284). Ideally, the thickness reading line should be a straight
line, indicating that the instrument is making the same thickness measurement
from the same reflector for each focal law (when that reflector is peaked up). In
this example, the uncalibrated line should look like the one shown in
Figure 17-10
on page 288.