910-269-EN, Rev. B, June 2011
Managing Special Waveform Functions (Conventional UT Mode) 107
Figure 5-4 The x-axis grid modes
4. Move the focus to the Y-Axis grid Mode parameter.
5. Choose the desired y-axis grid mode using the [RIGHT] and [LEFT] arrow keys,
or using the corresponding direct-access [P<n>] key (see
Figure 5-5 on page 108).
Standard grid: traditional flaw detector view with 10
divisions equally spaced across the screen range, and
the numbers 1 - 10 appearing below each division.
Sound Path grid: displaying actual sound path
measurements at equally spaced increments along the
horizontal axis. This mode displays 5 divisions, each
labeled with its corresponding sound path value
(depending on the 1/5 > Basic > Range, 1/5 > Basic >
Delay, and 3/5 > Meas Setup > Units settings).
Leg grid: displaying vertical lines representing angle
beam inspection legs. This mode displays up to four (4)
divisions, labeled L1 to L4, which represent each half-
skip distance of a test piece during an angle beam
inspection. The spacing and number of divisions
displayed depends on the 1/5 > Basic > Range, 1/5 >
Basic > Delay, and 1/5 > TRIG > Thick (material
thickness) parameters.