910-269-EN, Rev. B, June 2011
Calibrating the EPOCH 1000 Series (Phased Array Mode) 295
Figure 17-17 The transducer coupled to the 0-mark of the IIW block
6. Use the [GATE] key to position gate 1 so that the first reflection from the arc of the
block is exceeding the gate threshold. This reflection should be close to 4
in.
(100
mm).
7. Adjust the gain so that the echo amplitude is at approximately 80 %.