910-269-EN, Rev. B, June 2011
Managing the Data Logger 191
3. Press [ESCAPE] to save the report configuration and exit the setup page.
4. Depending on your selection in step 2:
a) Ensure that the CompactFlash card is inserted in the CompactFlash port in
the battery compartment (see section
1.4.4 on page 32).
OR
b) Connect the USB storage device to one of the two USB host ports in the
computer connection compartment on the right side of the instrument.
5. Press [2
nd
F], [RECALL SETUP] (PRINT) to save the screen capture file to the
destination specified in step
2.
The PNG format screen capture files that include the entire visible screen are
saved under the following path and file name:
\Olympus-NDT\EPOCH1000\<instrument_s/n>\ScreenCapture<#>.png
10.6 Instrument Resets
The EPOCH 1000 Series allows you to reset its current settings to default values, when
necessary. The instrument reset parameters are available in the Resets setup page
accessible by choosing 5/5
> Files > Reset. Use the [UP] and [DOWN] arrow keys to
highlight the desired data set. There are four data set options:
Master/Database
Represents the entire data logger and the live parameters.
Parameters
Represents the live parameters only.
Datalogger
Represents acquired data only.
Probe Library
Represents the DGS/AVG probe library data.
Up to five parameter functions are available at the bottom of the setup page to
perform actions on the selected data set:
Restore
Restores a backed up database, loading a back up file located at the root of a USB
storage device attached to the instrument, and then turns the instrument off. The
restored data set is available when you restart the instrument.