910-269-EN, Rev. B, June 2011
Calibrating the EPOCH 1000 Series (Conventional UT Mode) 143
Figure 9-5 The Enter Value for Zero Cal dialog box
11. Adjust the value to match the known thickness of the gated indication (0.100 in. in
this example), and then choose Continue to continue to the second calibration
step (see
Figure 9-6 on page 144).
If for any reason you need to exit without acquiring calibration data, use the
[RIGHT] and [LEFT] arrow keys to highlight Cancel, and then press [CHECK].