910-269-EN, Rev. B, June 2011
152 Chapter 9
Figure 9-10 Example of a gated calibration signal
13. Once a steady reading is achieved, choose 1/5 > Auto CAL > CAL Velocity.
The screen freezes and the Enter Value for Velocity Cal dialog box appears.
Figure 9-11 The Enter Value for Velocity Cal dialog box
14. Adjust the value to match the known thickness of the gated indication (0.300 in. in
this example), and then choose Calculate to complete the calibration process.