Figure 11-24 Active AWS with D rating
11.4.3 Scanning Gain
AWS codes require that you enter a certain amount of scanning gain to the Ref B gain
value. This allows you to locate flaws that might be smaller or deeper in the test piece
than the reference flaw.
To add scanning gain
1. Use the [GAIN] key to adjust the scanning-gain value necessary to perform the
inspection as outlined by the AWS Code.
2. Choose 6/6 > AWS > Scan dB to toggle the scanning gain on and off as necessary.