910-269-EN, Rev. B, June 2011
Managing the Data Logger 183
• Leg indicators for both gates
• Any displayed peak memory envelopes or peak hold A-scans
• Software feature/option setups
10.3.4 File Review
The EPOCH 1000 Series allows you to review the full contents of saved information
within each file.
To access the full file content review
1. Choose 5/5 > Files > Open to open the Open setup page.
2. Highlight the desired file in the file name list.
3. Press the [RIGHT] arrow key to highlight the ID list within the highlighted file
name.
4. Use the [UP] and [DOWN] arrow keys to highlight the desired ID.
5. Press [CHECK].
The contents of the file for the highlighted ID appears in the lower-half of the
screen.
6. Turn the adjustment knob to scroll up and down within the lower-half of the
screen to review the entire file content.
10.3.5 Quick Calibration File Recall
The EPOCH 1000 Series allows you to quickly recall a calibration file without entering
the file review submenu. The quick calibration recall feature is accessed by pressing
[RECALL SETUP] on the keypad. A dialog box appears, showing a list of all .CAL
files stored on the EPOCH
1000 Series instrument (see Figure 10-4 on page 184).