112 Chapter 6
Min Depth
Used to adjust the minimum depth value which triggers a minimum depth alarm.
Status
Used to toggle the state (On and Off) of the gate.
6.2 Quickly Adjusting Basic Gate Parameters
You can perform basic gate adjustments using the [GATE] direct-access key.
To quickly adjust the position of a gate
1. Press the [GATE] direct-access key.
The focus moves to direct-access gate button on the right side of the main display.
The gate button shows the first available gate parameter (see
Figure 6-3 on
page 112).
Figure 6-3 The direct-access gate button
2. Edit the value using the adjustment knob or the [UP] and [DOWN] arrow keys.
When needed, press [CHECK] to toggle between the coarse and fine mode.
3. To select the same parameter for another active gate, press [GATE] until the
desired gate is selected.
The gate button successively presents the G1Start, GIFStart, and G2Start
parameters.
The [GATE] key only allows access to gates which are currently active. To activate
a gate, select 2/5 > Gate<n> > Status = On.
Direct-access gate button