910-269-EN, Rev. B, June 2011
Calibrating the EPOCH 1000 Series (Conventional UT Mode) 165
Figure 9-22 The Enter Value for Zero Cal dialog box
7. Adjust the value to match the known thickness of the gated indication (0.500 in. in
this example), and then choose Continue to continue to the second calibration
step.
If for any reason you need to exit without acquiring calibration data, use the
[RIGHT] or [LEFT] arrow keys to highlight the Cancel box, and then press
[CHECK].