EasyManua.ls Logo

Olympus EPOCH 1000 Series - Calibrating the Sensitivity (Gain) with a Zero Degree Transducer

Olympus EPOCH 1000 Series
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
910-269-EN, Rev. B, June 2011
286 Chapter 17
The gated region must capture the wedge delay indication for all focal laws. If the
wedge delay indication is out of the gated region for any focal law during
calibration, the instrument does not properly calculate wedge delay for that/those
focal laws.
17.3.3 Calibrating the Sensitivity (Gain) With a Zero Degree Transducer
Sensitivity (gain) can be calibrated using a discrete reflector, such as a side-drilled
hole or a back-wall reflection. In the following example, the small side-drilled hole of
the IIW block is used to calculate sensitivity (gain).
To calibrate the sensitivity with a zero degree transducer
1. Press [RANGE], and then enter an appropriate range for the calibration.
For this example, use 2 in. (50 mm).
2. Select 1/5 > PA CAL > CAL Mode = PA CAL.
3. Press [ANGLE], and then adjust the selected focal law (angle) to 0°.
4. Couple the transducer to the IIW block over the side-drilled hole.
5. Use the [GATE] key to position gate 1 to surround only the first reflection from
the side-drilled hole.
6. Adjust the gain so that the echo amplitude is at approximately 80 %.
Downloaded from ManualsNet.com search engine

Table of Contents

Related product manuals