192 Chapter 10
Backup
Backs up the selected data set by creating a backup file at the root of the USB
storage device connected to the instrument (BackupDB.sdf file for
Master/Database and BackupParameters.sdf file for Parameters).
Reset
Restores values to factory or user defined default settings (defined using Create)
for the selected data set, and then turns the instrument off. The reset data set is
available when you restart the instrument.
Create
Used to create user defined default settings with the current data set values.
These settings are restored when you choose Reset.
Factory
Resets values to factory default settings, regardless of user defined default
settings, and then turns the instrument off. The factory data set is available when
you restart the instrument.
10.7 Hard Instrument Reset
In some situations, standard instrument resets may not correct an error occurring in
the EPOCH
1000 Series’ software. On rare occasions, the instrument may get stuck
while starting up, which does not allow you to access standard instrument resets.
The EPOCH 1000 Series allows you to perform a hard reset on the instrument.
A hard reset clears the data logger and returns all settings to the factory default
values.
To perform a hard instrument reset
1. Create and save an empty text file named RESET.TXT to the root folder of a USB
storage device.
2. Plug the USB storage device into the EPOCH 1000 Series.
3. Turn the instrument on while this USB storage device is connected.