116 Chapter 6
Gate % Amp Measurement
In Edge mode only, this selection allows you to define the method for measuring
amplitude of the gated indication:
Highest Peak
Acquires the amplitude measurement of the indication from the highest peak
within the gate.
1stPeak
Acquires the amplitude measurement of the indication from the first peak
within the gate. The peak must break the threshold of the gate to be
measured.
The EPOCH 1000 Series does not acquire measurement readings unless the
indication of interest is within a gated region of the screen. You should be careful
to adjust the Start, Width, and Level of a measurement gate so that only the
indication of interest falls within the gated region, per the measurement mode
definitions above.
6.5 Viewing Measurement Readings
The EPOCH 1000 Series provides six customizable measurement reading boxes to
display acquired measurements. These measurement readings must be properly
defined to view the desired information from a given indication.
For more information on defining measurement reading boxes, and for a full list of
possible measurement readings, see section
3.3.3 on page 71.
6.6 Gate Tracking and Echo-to-Echo Measurements
The gate tracking feature of the EPOCH 1000 Series allows you to make echo-to-echo
measurements whenever this feature is required by an application. Echo-to-echo
measurements can be made between: Gate
2 – Gate 1, Gate 2 – IF Gate, and Gate 1 –
IF
Gate.