910-269-EN, Rev. B, June 2011
30 Chapter 1
The phased array transducer connector features two alignment pins to ensure proper
connection to any phased array probe (see
Figure 1-8 on page 29). Two spring-loaded
screws are employed to maintain proper contact of the phased array probe to this
connector, and should always be fully engaged before use.
To connect a phased array transducer
1. Align the two guide pins on the phased array transducer with the guide-pin holes
on the instrument connector (see
Figure 1-8 on page 29).
2. Securely fit the transducer onto the instrument connector.
3. Tighten both spring-loaded retention screws on the phased array transducer.
Phased Array Connection Cover
The phased array transducer connector is not sealed for environmental hazards.
When a phased array transducer is not connected, proper care should be taken to
protect this connector from dust, dirt, liquid, and other potentially damaging
substances.
For this purpose, the EPOCH 1000iR and EPOCH 1000i include a protective cover to
protect the phased array transducer connector when not in use. The protective cover
includes a gasket for environmental seal, and attaches over the phased array
transducer connector using two thumb screws (see
Figure 1-9 on page 31).
When not in use, the phased array connection cover attaches to the battery
compartment cover of the EPOCH
1000 Series via two thumb screws. This allows easy
transportation of this connection cover and reduces the risk of loss or damage.