910-269-EN, Rev. B, June 2011
EPOCH 1000 Series Software Features 73
Gate 2 Surface Distance –
X Value
Horizontal distance minus X-value
(distance from beam index point to front
of wedge) in gate 2. Used with Angle.
Gate 1 Minimum Depth Minimum depth in gate 1. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 2 Minimum Depth Minimum depth in gate 2. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 1 Maximum Depth Maximum depth in gate 1. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 2 Maximum Depth Maximum depth in gate 2. Resets on gate
adjustment and on most pulser/receiver
adjustments.
Gate 1 Current Amplitude Amplitude measurement in gate 1.
Displays as % of full-screen height (FSH).
Gate 2 Current Amplitude Amplitude measurement in gate 2.
Displays as % of full-screen height (FSH).
Gate 1 Max. Amplitude
Maximum amplitude in gate 1. Resets on
gate adjustment and on most
pulser/receiver adjustments.
Gate 2 Max. Amplitude
Maximum amplitude in gate 2. Resets on
gate adjustment and on most
pulser/receiver adjustments.
Gate 1 Min. Amplitude Minimum amplitude in gate 1. Resets on
gate adjustment and on most
pulser/receiver adjustments.
Gate 2 Min. Amplitude Minimum amplitude in gate 2. Resets on
gate adjustment and on most
pulser/receiver adjustments.
Table 22 Available measurement readings (continued)
Icon Measurement readings Description