Page #165 background image
CMU RF: Power vs. Time Measurement
1100.4903.12 4.47 E-10
Resolution
Bandwidth
Resolution Bandwidth defines the resolution bandwidth of the measurement filter.
The nominal resolution bandwidth is the 3-dB bandwidth of the Gaussian meas-
urement filter. From a list discrete bandwidths between 10 Hz and 1 MHz can be
selected. The frequencies in the list are given by 1x10
n
Hz, 2x10
n
Hz, 3x10
n
Hz,
5x10
n
Hz where n=1 to 5. In addition the value 1 MHz can be selected.
Resolution bandwidths for the analyzer and the Spectrum measurement can be set
independently (see also p. 4.49).
Remote control
[SENSe:]POWer:FREQuency:BANDwidth[:RESolution] <Bandwidth>
[SENSe:]POWer:FREQuency:BWIDth[:RESolution] <Bandwidth>
Grid
The Grid checkbox switches The grid in the graphical test diagram on or off.
Remote control

Table of Contents