Audio Generator and Analyzer (with Option CMU-B41) CMU
1100.4903.12 6.70 E-10
CONFigure:AFANalyzer[:PRIMary]:CONTrol:DISTortion[:FREQuency] <Frequency> Frequency
CONFigure:AFANalyzer:SECondary:CONTrol:DISTortion[:FREQuency] <Frequency>
<Frequency>
Description of parameters Def. value Def. unit FW vers.
20 Hz to 20000
Hz
Reference frequency for distortion measurement 1000 Hz V2.10
Description of command
This command determines the reference frequency for the harmonic distortion measurement.
CONFigure:AFANalyzer[:PRIMary]:CONTrol:COUPling <Coupling> AF Path Coupling
CONFigure:AFANalyzer:SECondary:CONTrol:COUPling <Coupling>
<Coupling>
Description of parameters Def. value Def. unit FW vers.
AC |
DC
AC coupling of AF path
DC coupling of AF path
AC ([:PRIMary] channel)
DC (:SECondary channel)
– V3.05
Description of command
These commands determine the AF path coupling for measurements using the AF analyzer. In firmware versions
<3.05, the commands are replaced by two equivalent CONFigure:...COUPling commands.
Subsystem AFANalyzer...:FILTer (Filter)
The subsystem AFANalyzer:...FILTer configures the input path of the AF analyzer. The subsys-
tem corresponds to the Filter tab in the Analyzer Configuration menu. The input path of the AF analyzer
is as shown below:
Weighting
filter
Variable
band pass
AC Voltage (Peak) 1
AC Voltage (RMS) 1
AC Voltage (Peak) 2
AC Voltage (RMS) 2
Signal from
A
F IN (
[:PRIMary]
AF analyzer)
A
UX 1 (
:SECondary
AF analyzer)
Fixed
band pass
Fig. 6-1 AF analyzer input path configuration
CONFigure:AFANalyzer[:PRIMary]:FILTer:VBPass:CFRequency <Center> Frequency
CONFigure:AFANalyzer:SECondary:FILTer:VBPass:CFRequency <Center>
<Center>
Description of parameters Def. value Def. unit FW vers.
20 Hz to 20000 Hz Center frequency of band pass 1000 Hz V2.12
Description of command
This command determines the center frequency of the variable band pass.