CMU CMU Base System
1100.4903.12 6.15 E-10
Subsystem GTRMode (Local to Remote Switchover)
The GTRMode subsystem determines the behavior of the CMU in a local to remote transition. The set-
tings are provided in the Remote tab of the Setup menu.
SYSTem:GTRMode:COMPatible <Enable> Local to Remote
<Enable>
Parameter description Def. value Default unit FW vers.
ON |
OFF
Connection or call dropped on local to remote switchover/
generator switched off
Current signalling state or generator state maintained
ON – V3.05
Command description
This command defines the behavior of the CMU in a local to remote transition. The command is valid for all func-
tion groups and test modes, however, its effect depends on the test mode (Signalling or Non Signalling tests):
Signalling In the ON setting, the connection or call is dropped and the CMU returns to its default signal-
ling state (e.g. SOFF in the GSMxxx-MS function groups). In the OFF setting, all signalling
states are maintained. This makes it possible to switch the instrument to remote control with-
out dropping a call or connection. See also operating manuals for the different network tests.
Non Signalling In the ON setting, all generators are switched off. In the OFF setting, the current operating
state of all generators is maintained.
*GTL Go to Local
Command description
This command has no query form. It switches the instrument immediately back to local mode (manual control).
The CMU opens the menu of the current running measurement; see also section Return to Manual Control in
Chapter 5. This command can be used instead of the addressed command GTL is a serial connection is used.
Subsystem MQUeue
The MQUeue subsystem contains the commands for symbolic measurement ready evaluation. These
commands are used to query the contents of the measurement queue. To activate the measurement
queue, event reporting must not be switched off; see section Event Reporting in chapter 5.
SYSTem:MQUeue[:COMPlete][:LIST]? Complete Measurement Queue
Response
Parameter description Def. value Default unit FW vers.
<FGroup_Mode>,
<MeasObject>, ...
Function group and test mode
Ready measurement
"NONE",
"NONE"
– V2.10
Command description
This command is always a query. It returns and deletes the contents of the complete measurement queue. The
individual results are returned in chronological order (first in first out). Examples of possible responses are:
"RF_NSig","POWer","RF_NSig","SPECtrum"
Power and spectrum measurement ready in function group RF,
Test mode Non Signalling
"NONE", "NONE"
No ready measurements reported