Page #407 background imageLoading...
Page #407 background image
CMU Error Messages
1100.4903.12 9.1 E-4
9 Error Messages
SCPI Error Messages
SCPI error messages are the same in all SCPI instruments. The errors are assigned negative numbers.
The standard text of the error message is often supplemented by a comment from the CMU which
provides more detailed information (device-dependent information). Since this part depends on the
individual situation, it often contains more relevant information than the standard text.
No error
Error code Explanation
0
No error
This message is output when there are no entries in the error queue.
Command error
The following errors cause bit 5 in the ESR register to be set.
Error code Explanation
-100
Command error
Generic error message that cannot detect a more specific error.
-101
Invalid character
The command contains a character which is invalid for that type.
-102
Syntax error
The data type received is not accepted at this position.
-103
Invalid separator
The semicolon was omitted after a program message unit.
-104
Data type error
The recognized data element is of the wrong type (e.g. character data instead of numeric data)
-105
GET not allowed
A GET was received within a program message.
-108
Parameter not allowed
The command contains parameters at a position where they are not accepted.
-109
Missing parameter
The command does not contain the required parameters.

Table of Contents

Other manuals for Rohde & Schwarz CMU 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU 200 and is the answer not in the manual?

Rohde & Schwarz CMU 200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU 200
CategoryTest Equipment
LanguageEnglish

Related product manuals