CMU Measurement Control 
1100.4903.12 5.31  E-8 
Overlapping execution  CONTinue  is implemented as an overlapping command like INITiate. 
As a consequence, *OPC can not be used together with CONTinue.  
Stepping mode 
The stepping mode determines whether a measurement in the counting or 
continuous mode (see section Measurement Statistics on page 5.34) is 
interrupted after each evaluation period (or each statistics cycle, if cycles 
comprising several evaluation periods are defined) or not. The mode is set 
via the <Stepmode> parameter of the CONFig-
ure:<meas_obj>:CONTrol:REPetition CONTinuous |  1 ... 
10000, <StopCondition>, <Stepmode> command: 
<Stepmode> = STEP  The measurement is interrupted (⇒ measure-
ment state STEP) after each evaluation period, 
and the event reporting system (see p. 5.31) is 
invoked. The next measurement cycle must be 
started with the CONTinue:<meas_object> 
command. 
<Stepmode> = NONE  The measurement runs according to its repetition 
mode. Event reporting is invoked only when the 
measurement stops (status = RDY). 
Note:  STEP can be set in all repetition modes (single shot, continu-
ous, counting). For a single shot measurement which is always 
stopped after one evaluation period, the stepping mode has no 
effect. 
 
In function group RF Non Signalling, <meas_obj> can stand for any of the measurement objects POWer 
and SPECtrum. 
 
 
 
Event Reporting  
The event reporting system specifies in which way the CMU reports that a measurement or a meas-
urement step has been correctly terminated, i.e., that the measurement status STEP or RDY has been 
reached. Event reporting is configured for each measurement group individually by means of the com-
mands 
CONFigure:<meas_obj>:EREPorting SRQ | SOPC | SRSQ | OFF 
The parameters have the following meaning: 
SRQ  Service request. A service request is generated (i.e. bit no. 6 (RQS/MSS) of the status byte 
(STB) is set) whenever the measurement status STEP or RDY is reached (see section Ser-
vice Request on page 5.24). 
SOPC  Single operation complete. The operation complete bit (bit no. 0 in the event status register) 
is set whenever the measurement status STEP or RDY is reached (see section Status Re-
porting System on page 5.16). 
SRSQ  A service request is generated and the operation complete bit is set.  
OFF  No special action is taken when the measurement status STEP or RDY is reached. 
Note:  No action is taken if the STOP state is reached due to an explicit STOP command. 
 
 
Symbolic Measurement Ready Evaluation: 
If event reporting is enabled (i.e. CONFigure:<meas_obj>:EREPorting is not set to OFF), each 
measurement that reaches the STEP or RDY status causes an entry in the measurement queue. The 
measurement queue can be queried by means of the SYSTem:MQUeue[:COMPlete]<spec>? com-
mands described in Chapter 6.