CMU  Audio Generator and Analyzer (with Option CMU-B41) 
1100.4903.12 6.69  E-10 
CONFigure:AFANalyzer[:PRIMary]:MTReduce <Mode>, <Frequency>  Reduce Measurement Time
CONFigure:AFANalyzer:SECondary:MTReduce 
<Mode> 
Description of parameters  Def. value  Def. unit   
LOWF |  
 
EXPF 
Measurement time according to lowest frequency (10 
kHz) 
Measurement time according to <Frequency> 
LOWF   –    
<Frequency> 
Description of parameters  Def. value  Def. unit  FW vers. 
10 Hz to 20 kHz  Expected frequency of the audio input signal  100  Hz  V3.0 
Description of command 
This command is to reduce the measurement time the audio analyzer uses to calculate the measurement results 
(<Mode>=EXPF); see section Subsystem AFANalyzer... (Measured Values) on page 6.73. The measurement 
time must be adapted to the period of the input signal, which is the reverse of the frequency. Therefore, the audio 
measurement can be accelerated for high-frequency signals. In the LOWF mode, <Frequency> is not taken into 
account. 
 
 
 
Subsystem AFANalyzer...:CONTrol (Control) 
The subsystem AFANalyzer:...CONTrol defines the scope of the audio analysis and sets the reference 
frequency for the distortion measurement. The settings are provided in the Control and Distortion tabs 
of the Analyzer Configuration popup menu. 
 
CONFigure:AFANalyzer[:PRIMary]:CONTrol:REPetition  Test Cycles
CONFigure:AFANalyzer:SECondary:CONTrol:REPetition  
  <Repetition>, <StopCondition>, <Stepmode> 
<Repetition> 
Description of parameters  Def. value  Def. unit   
CONTinuous | 
SINGleshot | 
1 to 10000 
Continuous measurement (until STOP or ABORt) 
Single shot measurement (until Status = RDY) 
Multiple measurement (counting, until Status = 
STEP | RDY) 
SING –   
<StopCondition> 
Description of parameters  Def. value  Def. unit   
NONE       Continue measurement even in case of error  NONE –   
<Stepmode> 
Description of parameters  Def. value  Def. unit  FW vers. 
STEP | 
NONE 
Interrupt measurement after each statistic cycle Con-
tinue measurement according to its rep. mode 
NONE –  V2.10 
Description of command 
This command determines the number of statistics cycles and the stepping mode for the measurement. A stop 
condition is not available. 
Note:  In the case of READ commands (READ: to) the <Repetition> parameter has no effect; the measure-
ment is always stopped after a single shot.