Connecting the CMU to the Test Setup  CMU 
1100.4903.12 1.14  E-10 
Connecting the CMU to the Test Setup 
 
 
 
Warning: 
Connect external devices and peripherals only when the instrument is 
switched off or in STANDBY mode. Otherwise, future errors cannot be 
excluded. 
 
 
 
 
 
Connecting a Controller 
The CMU can be connected to an external controller via the GPIB bus (IEEE bus according to standard 
IEEE 488; throughout this documentation we will primarily use the term GPIB bus which is also used in 
the operating menus and in the SCPI command syntax) or via serial interface: 
Connection via GPIB 
bus 
 
 
 
IEEE 488
625
 
 
The CMU is connected to the GPIB interface of the controller via the GPIB 
bus connector (IEEE 488 / IEC 625) at the rear of the instrument and a 
shielded cable. The technical specifications of the GPIB interface are listed in 
section "Hardware Interfaces " in Chapter 8. 
 
 
 
GPIB Bus 
Configuration 
The parameters for GPIB bus control of the CMU are set in the Remote tab 
of the Setup popup menu (in the following abbreviated by Setup – Remote, 
see also chapter 4, Settings for Remote Control). 
!  To open the Setup - Remote menu, press the SETUP key at the front of 
the instrument and activate the Remote hotkey at the lower edge of the 
screen. 
!  Use the rotary knob to move the focus onto the SCPI Connection section 
of the Setup table. If necessary, press the rotary knob or the ON/OFF 
key to expand the parameters in the table (see Chapter 3). 
! In the Port table row select the GPIB bus interface for transmission. 
The bus address is factory-set to 20. It can be changed in the Primary 
Address input field.