CMU Error Messages 
1100.4903.12 9.5  E-4 
Device-specific error 
The following errors cause bit 3 in the ESR register to be set.  
 
Error code  Explanation 
-300 
Device-specific error 
-310 
System error 
An unspecified system error has occurred.  
-311 
Memory error 
An error was detected in the device's memory.  
-313 
Calibration memory lost 
Nonvolatile calibration data have been lost.  
-314 
Save/recall memory lost 
Nonvolatile saved data have been lost.  
-315 
Configuration memory lost 
Nonvolatile configuration data saved by the device have been lost.  
-330 
Self-test failed 
An error occurred during the internal self test.  
-350 
Queue overflow 
Error code entered in the queue in lieu of the code when the queue is full. It indicates that an error occurred 
but was not recorded in the queue. The original error message is lost. 
-360 
Communication error 
An unspecified communication error was detected.  
-361 
Parity error in program message 
Parity bit was not correct when data were received on a serial port.  
-362 
Framing error in program message 
No stop bit was detected when data were received on a serial port.  
-363 
Input buffer overrun 
Software or hardware input buffer on serial port overflows with data caused by improper or nonexistent 
pacing.