CMU Base System CMU
1100.4903.12 6.24 E-10
[SENSe:]SYNChronize:FREQuency:REFerence:LOCKed? Ref. Frequency Not Locked
Response
Parameter description Def. value Def. unit FW vers.
ON |
OFF
Synchronization to reference frequency achieved
Synchronization to reference frequency failed
– – V3.10
Command description
This command is always a query. It indicates whether the CMU is synchronized to the (external) reference fre-
quency.
Note: After activating the external reference frequency (command CONFigure:SYNChronize
:FREQuency:REFerence:MODE EXTernal) it is necessary to allow for a setting time
(~1 s) until the CMU has synchronized. In this case it is recommended to check whether
the reference frequency is locked before starting a measurement in remote control mode.