RF Measurements CMU
1100.4903.12 6.56 E-10
READ:SUBarrays:SPECtrum[:CURRent]? Subarray Results
READ:SUBarrays:SPECtrum:AVERage?
READ:SUBarrays:SPECtrum:MAXimum?
READ:SUBarrays:SPECtrum:MINimum? Start single shot meas. and return results ⇒ RUN
FETCh:SUBarrays:SPECtrum[:CURRent]?
FETCh:SUBarrays:SPECtrum:AVERage?
FETCh:SUBarrays:SPECtrum:MAXimum?
FETCh:SUBarrays:SPECtrum:MINimum? Read meas. results (unsynchronized) ⇒ RUN
SAMPle:SUBarrays:SPECtrum[:CURRent]?
SAMPle:SUBarrays:SPECtrum:AVERage?
SAMPle:SUBarrays:SPECtrum:MAXimum?
SAMPle:SUBarrays:SPECtrum:MINimum? Read results (synchronized) ⇒ RUN
Ret. values by subrange
Description of parameters Def. value Def. unit FW vers.
–128.0 dBm to + 48.0 dBm
...
–128.0 dBm to + 48.0 dBm
Power[1], 1
st
value for power
...
Power[x], xth value for power
NAN
...
NAN
dBm
...
dBm
3.0
Description of command
These commands are always queries. They return the power versus frequency in the subranges defined by
means of the CONFigure:SUBarrays:SPECtrum command. In the default setting of the configuration com-
mand the READ:SUBarrays..., FETCh:SUBarrays..., and SAMPle:SUBarrays... command group
is equivalent to the READ:ARRay..., FETCh:ARRay..., and SAMPle:ARRay... command group de-
scribed above.
The CONFigure:SUBarrays:SPECtrum command defines a maximum of 32 subranges. If one of the statisti-
cal modes (ARIThmetical, MINimum, MAXimum) or IVAL is set, only one value is returned by subrange.
The calculation of CURRent, AVERage, MINimum, and MAXimum results is explained in Chapter 3 (see Dis-
play Mode).