CMU Audio Generator and Analyzer (with Option CMU-B41)
1100.4903.12 6.67 E-10
[SENSe:]AFLevel[:PRIMary]:MAXimum <Level> AF Max. Level
[SENSe:]AFLevel:SECondary:MAXimum <Level>
<Level>
Description of parameters Def. value Def. unit FW vers.
0 V to +30 V Maximum audio input voltage 1 V V3.05
Description of command
This command defines the maximum expected AF input level. For firmware versions <3.05, only one
[SENSe:]AFLevel:MAXimum command is available to configure all audio channels.
[SENSe:]AFLevel:DEFault Default Settings
<Enable>
Description of parameters Def. value Def. unit FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.0
Description of command
If used as a setting command with the parameter ON this command sets all parameters of the subsystem to their
default values (the setting OFF results in an error message).