CMU Hardware Interfaces
1100.4903.12 8.3 E-8
2. Control bus with 5 lines
IFC (Interface Clear),
active LOW resets the interfaces of the instruments connected to the default setting.
ATN (Attention),
active LOW signals the transmission of interface messages
inactive HIGH signals the transmission of device messages.
SRQ (Service Request),
active LOW enables the connected device to send a service request to the controller.
REN (Remote Enable),
active LOW permits switchover to remote control.
EOI (End or Identify),
has two functions in connection with ATN:
ATN=HIGH active LOW marks the end of data transmission.
ATN=LOW active LOW triggers a parallel poll.
3. Handshake bus with three lines
DAV (Data Valid),
active LOW signals a valid data byte on the data bus.
NRFD (Not Ready For Data),
active LOW signals that one of the connected devices is not ready for data transfer.
NDAC (Not Data Accepted),
active LOW signals that the instrument connected is accepting the data on the data bus.

Table of Contents

Other manuals for Rohde & Schwarz CMU 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU 200 and is the answer not in the manual?

Rohde & Schwarz CMU 200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU 200
CategoryTest Equipment
LanguageEnglish

Related product manuals