Starting up the instrument C-3
17. Open the source and source side lock isolation valves.
18. Install the inner source chamber assembly and, if required, the solids
sample introduction probe.
See “Inserting the inner source chamber assembly into the outer source”
on page 7-33 and “Inserting the solids sample introduction probe in the
ion source” on page 6-9.
19. Open the GC interface reference inlet and pump valves.
See “The GC interface” on page 6-2.
20. Open the source lock pump valve.
21. Switch on the following MCBs:
Analyser Fine
Rotary Pumps:
Analyser
–Inlets 1
Source
Source Fine
22. Using the switch on each fore pump, switch on the fore pumps.
23. Allow the fore pumps to warm up for 5 minutes.
24. If applicable, open the Speedivalve on each fore pump, by turning it fully
counterclockwise, click Vacuum > Monitor.