EasyManua.ls Logo

Waters Autospec Premier - Page 58

Waters Autospec Premier
354 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-16 Tuning the instrument
To adjust the source slit for high resolution operation
1. In the Tune window’s Control 2 tab, set Detector (V) to 250 V.
2. On the Control 1 tab, set Collector Slit to 100.
Result: The peak becomes flat-topped, because the collector slit is wider
than the ion beam.
3. Note the maximum intensity of the peak, typically between 1 and 5 V,
depending on the level of PFK in the system.
Flat-topped mass 293 peak, intensity 1.0 V
4. On the Control 1 tab, adjust the Source Slit value so that the peak
intensity drops to 10% of the value obtained in step 3
Rationale: In going from 1000 to 10000 resolution, the ion beam
intensity drops to approximately 10% of its original value. This value is
the theoretical maximum transmission for the higher resolution, but can
be adversely affected by source conditions, such as the presence of a GC
column.

Table of Contents

Related product manuals