Preferences More Settings
Save and Restore Factory Default Gates and Threshold
You may find after using the Accutome A-Scan Plus for a
while, that you frequently have to adjust one or several
gates and threshold. If you continually make the same
adjustments to gates/threshold, you may want to save
the adjusted settings as the default for gates/threshold.
When you save gates/threshold the current locations of
the gates and threshold become the new patient default
locations.
To save gate/threshold settings as the default:
1. After you made gate adjustments to the current
waveform in the Measure Screen, return to the
Setup Screen (press the Preferences button, press
more settings...). Refer to
"How to Adjust Gates/
Threshold" on page 168 of this manual for more
information.
2. Press the button beneath the selection "Save
Gates". The Accutome A-Scan Plus will cue with a
message displayed at the top of the screen indicat
-
ing you have saved the current location of the
gates and threshold as the default.
To restore the locations of the gate and threshold to the
factory default:
1. Press and hold Save Gates button for approxi-
mately 1 sec. You will be prompted if you want to
restore gates/threshold to the factory default.