Accutome A-Scan Plus User’s Guide
Electrical Hazard and Safety
The instrument, foot switch and probe of the Accutome
A-Scan Plus are electrical/electronic devices.
Reasonable care should be taken when making
electrical connection and handling electrically powered
devices. Avoid use of damaged electrical equipment or
frayed electrical cords. If repair or maintenance is being
performed on the Accutome, the equipment must be
turned off and the power cord disconnected.
Covers must not be removed except by qualified
personnel. To avoid injury, do not operate the Accutome
without protective covers.
The system is intended to operate from a power source
of 100 - 240 VAC, 50/60 Hz. Protective earth grounding
through the grounding connection in the power cords is
essential for safe system operations.
Use only hospital-grade power cords and the connectors
supplied with the Accutome A-Scan Plus. Also, be sure
the power cords and connectors are in good condition.
To comply with IEC 601-1, the device should be
connected to earth via the ground terminal located next
to the DC input connector. Upon request, a ground cable
can be provided; Accutome part number: 24-4045.
Avoiding Equipment Damage
Attach the Accutome A-Scan Plus only to medical grade
IEC-601-1 Medically Approved equipment.
The Accutome A-Scan Plus provides no explosion
protection from static discharge or arcing components.
Do not operate the instrument in the presence of
explosive gases such as flammable mixtures of
anesthetic and air, or nitrous oxide.
To prevent overheating, keep ventilation holes, located
on the back of the instrument, free from obstruction.