Accutome A-Scan Plus User’s Guide
Contact Lens Method
The Contact Lens method requires you to enter the
following fields:
Kpre-the average K value before refractive
surgery
Current refraction in sphere and cylinder.
Refraction with contact lens on in sphere and
cylinder.
Base Curve (of the contact lens)
Power (of the contact lens)
1. Rotate the knob to select each field and press the
knob or the
button to activate the field. When
the field is active either rotate the knob to change
the value or manually enter the value with the key
-
board.
Figure 142 Post Rx Surgery Calculation with Contact Lens Method