268
Accutome A-Scan Plus User’s Guide
Acoustic
Output
Table 11 below provides the acoustic output reporting for
the following:
Transducer Model: Accutome A-Scan
Operating Mode: A-Scan
Application(s): Ophthalmic
Table 11 Acoustic Output Reporting Table for Track 1
Non-Autoscanning Mode
Acoustic
Output
MI
I
SPTA.3
(µW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum Value 0.14 2.62 4.11
Associ-
ated
Acoustic
Parame-
ter
P
r.3
(MPa)
0.41 ----- -----
W
o
(mW)
----- 0.198 0.198
f
c
(MHz)
9.34 9.34 9.34
Z
sp
(cm)
0.50 0.50 0.50
Beam
Dimen-
sions
x
-6
(cm) ----- 0.246 0.246
y
-6
(cm) ----- 0.332 0.332
PD
(µsec)
0.11 ----- 0.11
PRF
(Hz)
5.8 ----- 5.8
EBD Az. (cm) ----- 0.35 -----
Ele. (cm) ----- 0.35 -----

Table of Contents