Chapter 9 Reflection Operation
Performing a limited number of measurements
154 9102 Handheld Spectrum Analyzer Software version 5.31
Performing a limited number of measurements
The 9102 can run measurements continuously or a defined number of
times. Limiting the number of measurements can be useful for statis-
tical analyses.
1 In the main menu, select
Freq > Sweep.
The Sweep menu appears.
2 Select the trigger mode: Push
Cont. for continuous measurements
or
Single for a limited number of measurements.
The selected trigger mode is highlighted.
3 To enter the number of measurements, push the
Single Count
softkey, enter a number in the range from 1 to 1000 and press
E
NTER.
If Trigger Mode is set to Single, the 9102 performs the defined
number of measurements and enters hold mode.
– To restart single-mode measurements, push the H
OLD/RUN func-
tion key or the
Single softkey.
– To stop a continuous measurement, push the H
OLD/RUN function
key. Push it again to resume measurements.
Setting up the trace
The trace functions provide different views of the measurements, for
example the actual measurement or an average over the last couple of
measurements. You can even select two different views of the
measurement. Another possibility is to compare the actual measure-
ment with an older measurement which has been stored in the 9102
and loaded to one of the trace views (see section “Storing and loading
traces” on page 158).
Selecting the trace
mode
The 9102 has five different modes to display a trace:
– In actual mode, the 9102 shows a complete new measurement in
each trace. Subsequent traces are independent of each other.
– In hold mode, the last measurement is kept on the display;
measurements continue but are not displayed.
– In max hold mode, the 9102 takes new measurements and, for
each frequency point, compares the new measurement with the
previous result. If the new measurement value is higher than the
previous result, the new measurement value becomes the new
result value; otherwise the old result value is kept. This way, the
highest result since the start of the Max hold measurement (or a
parameter change) is kept and displayed.
– Similarly, in min hold mode, the 9102 takes new measurements
and compares the new measurement with the previous result. If the
new measurement value is lower than the previous result, the new
measurement value becomes the new result value; otherwise the