Chapter 2 General Operation
Working with the markers
9102 Handheld Spectrum Analyzer Software version 5.31
33
2 Select the softkey for the delta marker that you want to disable (e.g.
DA2). If you use more than four markers you can display the soft-
keys for marker 5 and 6 by pressing the
Mkr More softkey.
3 To disable the marker completely, push the softkey until “OFF“ is
highlighted.
To turn the delta marker into a normal marker displaying absolute
values again toggle the softkey to “ON“.
Setting a marker on a
frequency relative to
marker 1
1 Enable a delta marker as described in “Enabling a delta marker”.
For the selected delta marker, the input field shows the frequency
relative to marker 1.
2 Enter a (signed) frequency relative to the frequency at marker 1,
either using the numeric keys and the respective enter key, or
moving the cursor to that frequency with the help of the cursor
keys.
The marker field indicates the desired frequency offset for that
marker, along with the power level relative to the power at marker
1.
Changing the center
frequency with a marker
This function modifies the center frequency, adapting the frequency of
a selectable marker.
1 From the main menu, select
Marker, or push the MKR function key
in any menu.
The Marker menu and the input field for marker 1 appear.
2 If you want to use the frequency at a marker position other than
marker 1, push the appropriate softkey (
A2 through A6 or B2
through
B6). If you use more than four markers you can display the
softkeys for marker 5 and 6 by pressing the
Mkr More softkey.
3 Push the
Marker to Center softkey.
The center frequency changes to the frequency at which the
selected marker is located. The frequency span only changes if the
change in center frequency would lead to an invalid start or stop
frequency.
Changing the reference
level with the marker
The reference level can be changed to the level at a marker position as
follows:
1 From the main menu, select
Marker, or push the MKR function key
in any menu.
2 Select the
Mkr Transfer menu.
3 If you want to use the power level at a marker position other than
marker 1, push the appropriate softkey (
A2 through A6 or B2
through
B6). If you use more than four markers you can display the
softkeys for marker 5 and 6 by pressing the
Mkr More softkey.

Table of Contents