Chapter 16 SCPI Command Reference
Instrument commands
9102 Handheld Spectrum Analyzer Software version 5.31
311
:MMEMory:DELete:AFACtor:ALL
:MMEMory:DELete:CFACtor
:MMEMory:DELete:CFACtor:AL
L
Instrument commands
:INSTrument:SELect
Syntax
:MMEMory:DELete:AFACtor:ALL
Parameters
There are no parameters.
Command
Deletes all the files in the antenna factor directory on the flash disk.
Query
There is no query form of this command available.
Example
:MMEMory:DELete:AFACtor:ALL
Syntax
:MMEMory:DELete:CFACtor <string1>
Parameters
string1 is a string (text) parameter. The maximum length of string1 is
11 characters.
Command
Deletes file string1 in the cable factor directory on the flash disk. Files in
this directory contain cable factor compensation settings on the 9102.
Query
Reads and returns the name of the file deleted last with this command.
Example
:MMEMory:DELete:CFACtor "CFAC7"
Syntax
:MMEMory:DELete:CFACtor:ALL
Parameters
There are no parameters.
Command
Deletes all the files in the cable factor directory on the flash disk.
Query
There is no query form of this command available.
Example
:MMEMory:DELete:AFACtor:ALL
Syntax
:INSTrument:SELect <PredefExpr>
Parameters
PredefExpr is one of the following expressions:
SANalyzer|CPOWer|TRANsmission|DTF|SGENerator|
RFLection|CLOSs|EMF|PSENsor
.
Default is SANalyzer.
Command
Selects the measurement mode. Available modes are spectrum analysis
channel power, transmission, distance to fault, signal generator, reflection,
cable loss, electro magnetic field and power sensor.
Selects the measurement mode. Available modes are spectrum analysis,
channel power, transmission, distance to fault, signal generator, reflection,
cable loss, electro magnetic field and power sensor.
Query
Reads and returns the current setting.