Chapter 16 SCPI Command Reference
SCPI errors
334 9102 Handheld Spectrum Analyzer Software version 5.31
:SERVice:EDEVice:TEXT?
:SERVice:EDEVice:CALibration:DATe?
:SERVice:EDEVice:CALibration:NUMBer?
SCPI errors
The following table lists the error numbers which the 9102 may return
in case of a problem.
Syntax
:SERVice:EDEVice:TEXT?
Parameters
There are no parameters.
Command
There is only a query form of this command available.
Query
Reads and returns the description text of the external device connected to
your Aeroflex 9102. The command will return a string.
Example
:SERVice:EDEVice:TEXT?
String returned in this example: "VSWR Bridge"
.
Syntax
:SERVice:EDEVice:CALibration:DATe?
Parameters
There are no parameters.
Command
The command form is not available.
Query
Returns date of last calibration by Aeroflex of the external device con-
nected in the form yyyy,mm,dd.
Example
:SERVice:EDEVice:CALibration:DATe?
String returned in this example: "2004/12/31"
.
Syntax
:SERVice:EDEVice:CALibration:NUMBer?
Parameters
There are no parameters.
Command
There is only a query form of this command available.
Query
Reads and returns the calibration number of the external device connected
to your Aeroflex 9102. The command will return a string.
Example
:SERVice:EDEVice:CALibration:NUMBer?
String returned in this example: "1234"
.
Error
numbe
r
Error description
Command errors
-100 Internal error only, for debugging purposes (Command
error)
-101 Invalid character in command string