Chapter 16 SCPI Command Reference
System commands
9102 Handheld Spectrum Analyzer Software version 5.31
269
:SYSTem:DNAMe
:SYSTem:OPTions
:SYSTem:SCReendump:COLor?
Query
Returns the error codes of all unread error messages in the internal error
queue of the Aeroflex 9102. The string returned will contain a maximum of
100 integers, separated by commas.
Note: A list of error messages can be found in section “SCPI errors” on
page 334.
Example
:SYSTem:ERRor:CODE:ALL?
String returned in this example: "-113,-112,0,0,0,0,0,0,0,0"
This means that there were two unread error messages in the error queue.
Syntax
:SYSTem:DNAMe <string1>
Parameters
string1 is a string (text) parameter. The maximum length of string1 is
11 characters.
Command
Sets a user-definable device name for the 9102 to use this unique name on
protocols.
Query
Reads and returns the device name set on this 9102.
Example
:SYSTem:DNAMe "Develop5"
:SYST:DNAM?
String returned in this example: "Develop5"
Syntax
:SYSTem:OPTions <string1>
Parameters
string1 is a string (text) parameter. The maximum length of string1 is
25 characters.
Command
Enables or disables a software option on the Aeroflex 9102 with the appro-
priate option key string1.
Query
There is no query form available.
Example
:SYSTem:OPTions "00172663845433"
Enables the option corresponding to the key.
Syntax
:SYSTem:SCReendump:COLor? <int1>
Parameters
int1 is a integer in the range of 0 to 255.
Command
There is only the query form available.
Query
Returns the RGB color value (in hexadecimal) of the given color palette
entry. Max. 256 palette entries are available.
Example
:SYST:SCR:COL? 5
String returned in this example:
"a800a8"